Measuring fifteen endogenous estrogens simultaneously in human urine by high-performance liquid chromatography-mass spectrometry

  1. Author:
    Xu, X.
    Veenstra, T. D.
    Fox, S. D.
    Roman, J. M.
    Issaq, H. J.
    Falk, R.
    Saavedra, J. E.
    Keefer, L. K.
    Ziegler, R. G.
  2. Author Address

    NCI, Lab Proteom & Analyt Technol, SAIC Frederick Inc, Frederick, MD 21702 USA. NCI, Epidemiol & Biostat Program, Div Canc Epidemiol & Genet, Frederick, MD 21702 USA. NCI, Comparat Carcinogenesis Lab, Canc Res Ctr, Frederick, MD 21702 USA. NCI, Basic Res Program, SAIC Frederick Inc, Frederick, MD 21702 USA Veenstra, TD, NCI, Lab Proteom & Analyt Technol, SAIC Frederick Inc, Frederick, MD 21702 USA
    1. Year: 2005
    2. Date: OCT 15
  1. Journal: Analytical Chemistry
    1. 77
    2. 20
    3. Pages: 6646-6654
  2. Type of Article: Article
  1. Abstract:

    A sensitive, specific, accurate, and precise high-performance liquid chromatography-electrospray ionization-tandem mass spectrometry method for measuring the, absolute quantities of 15 endogenous estrogens and their. metabolites in human urine has been developed and validated. The method requires a single hydrolysis/ extraction/derivatization step and only 0.5 mL of urine, yet is capable of simultaneously quantifying estrone and its 2-, 4-methoxy and 2-, 4-, and 16 alpha-hydroxy derivatives, and 2-hydroxyestrone-3-methyl ether; estradiol and its 2-, 4-methoxy and 2-, 16 alpha-hydroxy derivatives, 16-epiestriol, 17-epiestriol, and 16-ketoestradiol in pre- and postmenopausal women as well as men. Standard curves are linear over a 10(3)-fold concentration range with the standard error of the estimate (SEE) and the relative standard error of the estimate (RSEE) for the linear regression line ranging from 0.0131 to 0.1760 and 1.2 to 7.3%, respectively. The lower limit of quantitation for each estrogen is 0.02 ng/0.5 mL urine sample (2 pg on column), with the percent recovery of a known added amount of compound (accuracy) of 96-107% and an overall precision, including the hydrolysis, extraction, and derivatization steps, of 1-5% relative standard deviation (RSD) for samples prepared concurrently and 1-12% RSD for samples prepared in separate batches. Since individual patterns of estrogen metabolism may influence the risk of breast cancer, accurate, precise, and specific measurement of endogenous estrogen metabolites in biological matrixes will facilitate future research on breast cancer prevention, screening, and treatment

    See More

External Sources

  1. WOS: 000232785900018

Library Notes

  1. No notes added.
NCI at FrederickClose Button

You are leaving a government website.

This external link provides additional information that is consistent with the intended purpose of this site. The government cannot attest to the accuracy of a non-federal site.

Linking to a non-federal site does not constitute an endorsement by this institution or any of its employees of the sponsors or the information and products presented on the site. You will be subject to the destination site's privacy policy when you follow the link.

ContinueCancel