Fast and accurate brain T-2 relaxation time quantification in animal models calibrated using gel phantoms and in vivo data suitable for imaging at a biosafety level 4 environment

  1. Author:
    Castro, Marcelo A.
    Laux, Joseph
    Friel, Harry
    Solomon,Jeffrey
    Byrum, Russell
    Rojas, Oscar
    Thomasson, David
    Bollinger, Laura
    Hammoud, Dima
    Lee, Ji Hyun
  2. Author Address

    NIAID, Integrated Res Facil, NIH, 8200 Res Pl, Frederick, MD 21702 USA.Philips Healthcare, 595 Miner Rd, Highland Hts, OH 44143 USA.NCI, Frederick Natl Lab Canc Res, Clin Monitoring Res Program Directorate, 8200 Res Pl, Frederick, MD 21702 USA.NIH, Clin Ctr, Ctr Infect Dis Imaging, 10 Ctr Dr, Bethesda, MD 20814 USA.
    1. Year of Conference: 2020
  1. Conference Name: MEDICAL IMAGING 2020: PHYSICS OF MEDICAL IMAGING
  2. SPIE-INT SOC OPTICAL ENGINEERING,
  3. Houston, TX
    1. 11312
    2. Pages: 1131233
  4. Type of Work: Proceedings Paper
  5. Article Number: ARTN 1131233
  6. ISBN: 1605-7422
  1. Abstract:

    The brain is an organ of interest in high-consequence infectious disease. Subtle blood-brain-barrier disruption and edema can be identified and quantified using T-2 relaxation time as a biomarker. Accuracy is required to correlate changes in imaging findings with biological processes and therapeutic effectiveness. Dual echo time Fluid- Attenuated Inversion Recovery images, which minimizes the partial volume effects from cerebrospinal fluid, can be used to quantify T-2 relaxation. This sequence also meets the acquisition time requirements of biosafety level-4 facilities, where cohorts of animals are imaged on a tight schedule until terminal stages, when the animals may not survive long scans. However, the echo time used to compute T-2 varies among slices due to k-space mapping in fast spin-echo sequence. Therefore, an effective echo time instead of the prescribed echo time should be considered. We hypothesize that accurate effective echo time can be estimated by optimization for each slice ordering. A longer and more accurate T-2-weighted spin-echo (multiecho) sequence to compute T-2 maps from mono-exponential fitting was validated using gel phantoms and used as ground truth. A brain of a non-human primate was imaged in vivo using both T-2 sequences. T-2 maps were generated using both methods and a correction factor was computed by linear fitting for each set of echo times and slice ordering. The coefficient of variation within phantoms and the Pearson's correlation coefficient between in vivo ground truth T-2 and each map were used to assess its accuracy and quality.

    See More

External Sources

  1. DOI: 10.1117/12.2548619
  2. WOS: 000671890600100

Library Notes

  1. Fiscal Year: FY2019-2020
NCI at Frederick

You are leaving a government website.

This external link provides additional information that is consistent with the intended purpose of this site. The government cannot attest to the accuracy of a non-federal site.

Linking to a non-federal site does not constitute an endorsement by this institution or any of its employees of the sponsors or the information and products presented on the site. You will be subject to the destination site's privacy policy when you follow the link.

ContinueCancel