Element of a validation method for MU-B3 monoclonal antibody using an imaging capillary isoelectric focusing system

  1. Author:
    Janini, G.
    Saptharishi, N.
    Waselus, M.
    Soman, G.
  2. Author Address

    NCI, SAIC Frederick Inc, Analyt Chem Lab, POB B,Bldg 469,Room 155, Ft Detrick, MD 21702 USA NCI, SAIC Frederick Inc, Analyt Chem Lab, Ft Detrick, MD 21702 USA NCI, SAIC Frederick Inc, Biopharmaceut Dev Program, Ft Detrick, MD 21702 USA Janini G NCI, SAIC Frederick Inc, Analyt Chem Lab, POB B,Bldg 469,Room 155, Ft Detrick, MD 21702 USA
    1. Year: 2002
  1. Journal: Electrophoresis
    1. 23
    2. 11
    3. Pages: 1605-1611
  2. Type of Article: Article
  1. Abstract:

    This paper presents an imaging capillary isoelectric focusing (CIEF) assay for the determination of the identity, stability, and isoform distribution of a murine monoclonal antibody (MU- B3). The experiments were conducted using a Convergent Bioscience iCE280 instrument. The optimum carrier ampholyte composition that gave the best peak separation was found to be 25% Pharmalyte pH 3-10 and 75% Pharmalyte pH 5-8. The antibody gave a highly reproducible CIEF profile with three major peaks having average isoelectric point (pl) values of 6.83, 6.99, and 7.11. Intraday and interday reproducibility of pl values was found to be within RSD of 0.5%. The CIEF profile was also the same, with an alternate column cartridge and alternate batches of methyl cellulose. A plot of peak areas versus MU-B3 concentration was linear (R-2 = 0.995) up to a concentration of 0.5 mg/mL in the sample solution. Peak area measurements were reproducible to within 7% RSD. The CIEF profiles of two other antibodies were distinctly different from the profile of MU-B3, showing that the assay is specific. After a sample of MU-B3 was subjected to heat stress by exposure to heat at 55degreesC for 4 h, its CIEF profile was altered with extra peaks appearing at lower pl values, indicating that the assay could be used to monitor stability. The result of the heat stress experiment was also confirmed with a parallel slab-gel IEF analysis of the antibody sample before,and after application of the heat stress. The results of this work suggest that imaging CIEF can be used for product testing under a quality control environment. The assay can be used for pl profiling of proteins and for monitoring structural changes (deamidation, glycosylation, etc.) during the manufacturing process and upon storage.

    See More

External Sources

  1. No sources found.

Library Notes

  1. No notes added.
NCI at Frederick

You are leaving a government website.

This external link provides additional information that is consistent with the intended purpose of this site. The government cannot attest to the accuracy of a non-federal site.

Linking to a non-federal site does not constitute an endorsement by this institution or any of its employees of the sponsors or the information and products presented on the site. You will be subject to the destination site's privacy policy when you follow the link.

ContinueCancel