    Falk, Roni T.; Ziegler, Regina G.] NCI, Epidemiol & Biostat Program, Div Canc Epidemiol & Genet, Bethesda, MD 20892 USA. [Xu, Xia, Veenstra, Timothy D.] SAIC Frederick Inc, Adv Technol Program, Lab Proteom & Analyt Technol, Frederick, MD USA. [Keefer, Larry] NCI, Comparat Carcinogenesis Lab, Ctr Canc Res, Frederick, MD 21701 USA.
    1. Year: 2008
  1. Journal: Cancer Epidemiology Biomarkers & Prevention
    1. 17
    2. 12
    3. Pages: 3411-3418
  2. Type of Article: Article
  1. Abstract:

    Background: Accurate, reproducible, and sensitive measurements of endogenous estrogen exposure and individual patterns of estrogen metabolism are needed for etiologic studies of breast cancer. We have developed a high-performance liquid chromatography-tandem mass spectrometry method to quantitate simultaneously 15 urinary estrogens and estrogen metabolites (EM): estrone, estradiol, 3 catechol estrogens, 5 estrogens in the 16 alpha pathway, including estriol, and 5 methoxy estrogens. Methods: Overnight urines were obtained from 45 participants. For the reproducibility study, two blinded, randomized aliquots from 5 follicular and 5 luteal premenopausal women, 5 naturally postmenopausal women, and 5 men were assayed in each of four batches. Assay coefficients of variation and intraclass correlation coefficients were calculated with ANOVA models. Data from the additional 25 participants were added to compare EM levels by menstrual/sex group and assess interindividual variability. Results: For each EM, overall coefficients of variation were <= 10%. Intraclass correlation coefficients for each menstrual/sex group were generally >= 98%. Although geometric mean EM concentrations differed among the four groups, rankings were similar, with estriol, 2-hydroxyestrone, estrone, estradiol, and 16-ketoestradiol accounting for 60% to 75% of total urinary EM. Within each group, interindividual differences in absolute concentrations were consistently high, the range was 10- to 100-fold for nearly all EM. Conclusion: Our high-performance liquid chromatography-tandem mass spectrometry method for measuring 15 urinary EM is highly reproducible, and the range of EM concentrations in each menstrual/sex group is quite large relative to assay variability. Whether these patterns persist in blood and target tissues awaits further development and application of this method. (Cancer Epidemiol Biomarkers Prev 2008,17(12):3411-8)

    See More

External Sources

  1. PMID: 19064556

Library Notes

  1. No notes added.
NCI at Frederick

You are leaving a government website.

This external link provides additional information that is consistent with the intended purpose of this site. The government cannot attest to the accuracy of a non-federal site.

Linking to a non-federal site does not constitute an endorsement by this institution or any of its employees of the sponsors or the information and products presented on the site. You will be subject to the destination site's privacy policy when you follow the link.

ContinueCancel